Manual
● Standards IEC61010-1 Measurement CAT.Ⅲ 300V, pollution degree 2 / Measurement CAT.Ⅱ 600V, pollution degree 2
IEC61010-031
IEC61326
● Mode of operation △∑ mode
● Indication LCD maximum value 6000 (ACV/A, DCV/A, Ω) / maximum value 9999 (Hz) / maximum valve 4000 (F) , units, symbols
● Over range display "OL" symbol is displayed on the LCD.
"OL" displayed when exceeding the measuring ranges. (except for AC/DC 600V ranges and AC/DC 10A ranges.)
● Auto-ranging Range shifts to upper range when indicated value is more than 6040.
Range shifts to lower range when indicated value is less than 560.
● Sampling rate approx. 400ms
● Operating Environmental −indoor use
conditions −altitude up to 2000m
● Temperature & Humidity range
(guaranteed accuracy) 23℃±5℃ Relative humidity: less than 75%
● Operating Temperature &
Humidity range 0℃∼+40℃ Relative humidity: less than 80%
● Storage Temperature &
Humidity range -20℃∼+60℃ Relative humidity: less than 90%
● Insulation Resistance It should be more than 100MΩ/DC1000V between electrical circuit and enclosures.
● Withstand Voltage It should be more than AC3700V/ for one minute between electrical circuit and enclosures.
● Overload Protection Voltage function : 720V (RMS.) 10seconds
Resistance function : 600V (RMS.) 10seconds
Capacitance function : 600V (RMS.) 10seconds
Frequency function : 600V (RMS.) 10seconds
Current function : uA, mA : Protected by 600V 0.8A fuse
A:Protected by 600V 10A fuse
● Dimensions / Weight approx. 161(L) x 82(W) x 50(D) mm / approx. 280g(including batteries, holster installation)
● Power source Two R6P(AA)1.5V or equivalent
● Accessories Test lead 1set / R6P (AA) 2p.c.s / Holster 1 / Instruction manual 1
● Fuse F600V/800mA (fast acting type), φ6.3 x 32mm / F600V/10A(fast acting type), φ6.3 x 32mm
CAUTION
● The voltage shown above is the overload protection(overvoltage protection)for the instrument. Make sure not to exceed the value of voltage shown above.
● In locations subject to radio frequency interference, the product may malfunction and it will recover automatically after the test.