Instruction Manual
89
4.3.10CurrentMeasurement(µA/mA/A)
W ARNIN G
To avoid damage to instrument or equipment
• Before starting measurement, make sure that the position of function switch and the
input terminals for connecting the test leads are appropriate for the desired mode of
measurement.
• Themaximuminputcurrent(limitedbyfuses)ofthe“µA”and“mA”rangesis440mA.
Be sure not to exceed the limit in the 500 mA range.
Be careful not to burn yourself
• When measuring more than 6A under exceeding40
℃
conditions, the continuous
measuringtimeshallbewithin3minutes,thenkeepdisconnectedformorethan10
minutes.
1) Turn the function switch to the “µA ” , “mA ” or “A ”position.
(If the magnitude of the current being measured is not known,
select the “A ”position. Make sure the current being measured
is no more than 440mA before the “µA ” or “mA ” position is
selected.)
2) Please select between DC and AC. When selecting AC, press
the SELECT key.
3)Plugtheblacktestleadintothe“COM”inputterminalandthe
red test lead into the “A ”input terminal.
If the current is in the order of mA or less, plug the red test lead
intothe“µA•mA”inputterminal.
4) Connect the test leads to the circuit under test and then read the
value when it stabilizes.
Test leads
Red
Black
Test leads
Red Black