Instruction Manual
113
6.2 Fuse Replacement
If a current greater than the rated value flows when the instrument is in the current-
measurement range, a protection fuse may blow.
If this happens, replace that fuse. The instrument contains the following types of fuses.
W ARNIN G
• TurnthefunctionswitchtoOFF(turnoffthepower).
• Be sureto disconnect the instrument fromthecircuit under test and testleads before
replacing the fuses.
• Donotoperatetheinstrumentwiththecasingleftopen.
• In order toavoid damageto the instrumentor any possibleaccident,use fuses ofthe
speciedrating.
Fuse rating: F1 M-8926
(440mA/1000V, SIBA GmbH & Co. KG, 50 210 06.0.44)
F2 M-8927 (10A/1000V, SIBA GmbH & Co. KG, 50 199 06.10)
High breaking capacity type
To replace the fuse:
1) Remove the screw on the back of the casing.
2) Remove the back cover.
3)Removetheblownfusefromthefuseholder.
4) Install a new fuse in the holder.
(Make sure the fuse rating.)
5) Close the casing and fasten it with the screw.
F1:
(440mA/1000V)
F2:
(10A/1000V)
Screw hole
<Rear View>
Back cover removed