取扱説明書 INSTRUCTION MANUAL デジタルクランプメータ DIGITAL CLAMP METER KEW SNAP SERIES MODEL 2 0 0 2 PA 2 0 0 2 R KYORITSU ELECTRICAL INSTRUMENTS WORKS,LTD.
ENGLISH Contents 1.Safety Warnings ……………………………………………………… 26 2.Features ……………………………………………………………… 28 3.Specifications 3−1 MODEL 2002PA………………………………………………… 29 3−2 MODEL 2002R ………………………………………………… 31 4.Instrument Layout …………………………………………………… 34 5.Preparation for Measurement 5−1 Battery Voltage Check ……………………………………… 35 5−2 Checking Switch Setting and Operation …………………… 35 6.Measurement 6−1 Current Measurement…………………………………………36 6−2 Voltage Measurement ……………………………………… 38 6−3 Resistance Measurement …………………………………… 39
1.Safety Warnings ○This instrument has been designed and tested according to IEC Publication 61010;Safety Requirements for Electronic Measuring Apparatus. This instruction manual contains warnings and safety rules which must be observed by the user to ensure safe operation of the instrument and to retain it in safe condition. Therefore, read through these operating instructions before starting using the instrument.
DANGER ●Never make measurement on the circuit above 750VAC or 1000VDC. ●Do not attempt to make measurement in the presence of flammable gasses, fumes, vapor or dust. Otherwise, the use of the instrument may cause sparking, which can lead to an explosion. ●Transformer jaw tips are designed not to short the circuit under test. If equipment under test has exposed conductive parts, however, extra precaution should be taken to minimize the possibility of shorting.
2.Features ●Tear-drop-shaped jaws for ease of use in crowded cable areas and other tight places. ●Accurate true-RMS reading of AC current or voltage with distorted waveform (MODEL2002R) ●Provides a wide measuring range from 0 up to 2000A. ●Terminal cover to avoid the use of an incorrect terminal. ●Measures current variation as short as 10 msec with peak-hold feature. ●Provides output to a chart recorder for current variation recording. ●Designed to international safety standards. IEC61010-1 (CAT.
3.Specifications 3−1 MODEL 2002PA ●Measuring Ranges and Accuracy(at 23±5℃, relative humidity 45-75%) AC Current 400A, 2000A ∼ ∼ Range Measuring Range Resolution 400A 0∼400.0A 0.1A 2000A 0∼1000A 1000∼1500A 1500∼2000A 1A AC Voltage( Accuracy(Frequency Range)Maximum Measurement Time ±1.0%rdg±3dgt(50/60Hz) Continuous ±2.0%rdg±3dgt(40∼1kHz) ±1.0%rdg±3dgt(50/60Hz) 15min ±3.0%rdg±3dgt(40∼1kHz) 5min ±3.0%rdg (50/60Hz) ∼V)Auto-ranging Range Measuring Range Resolution Accuracy(Frequency Range) 40V 0∼40.
●Operating System ●Display Dual Integration Liquid crystal display with a maximum count of 4000 ●Low Battery Warning "BATT" symbol is displayed on the digital display. ●Overrange Indication "OL" is displayed where input exceeds the upper limit of a range ●Response Time Approx. 2 seconds ●Sleep function Automatically powered down in about 10 minutes after the last switch operation ●Data Hold Available in all ranges provided the peak measurement mode is deactivated.
3−2 MODEL 2002R ●Measuring Ranges and Accuracy(at 23±5℃, relative humidity 45-75%) AC Current 400A, 2000A (9 counts or less is corrected to 0) ∼ ∼ Range Measuring Range Resolution 400A 0∼400.0A 0.1A 2000A 0∼1000A 1000∼1500A 1500∼2000A 1A Accuracy(Frequency Range)Maximum Measurement Time ±1.5%rdg±3dgt(45∼65Hz) ±2.5%rdg±3dgt(40∼1kHz) Continuous ±2.0%rdg±5dgt(45∼65Hz) 15min ±3.0%rdg±5dgt(40∼1kHz) 5min ±4.
●Operating System ●Display Dual Integration Liquid crystal display with a maximum count of 4000 ●Low Battery Warning "BATT" symbol is displayed on the digital display. ●Overrange Indication "OL" is displayed where input exceeds the upper limit of a range ●Response Time Approx. 2 seconds (at full scale) ●Sleep function Automatically powered down in about 10 minutes after the last switch operation ●Data Hold Available in all ranges provided the peak measurement mode is deactivated.
― 33 ―
4.Instrument Layout Transformer Jaw Jaw Trigger Function Switch Data Hold Switch Display Reset Switch Terminal Cover Mode Switch Drop Prevention Strap OUTPUT Terminal COM Terminal ︸ Test Leads Mー7107 V/Ω Terminal Recorder Output Plug Mー8102 ● LCD INDICATOR DC Current AC Current Data Hold Mode Resistance Negative Potential Voltage Current Continuity Check Mode Battery Voltage Alert ― 34 ― Peak Mode Measurement
5.Preparation for Measurement 5−1 Checking Battery Voltage ①Set the function selector switch to any position other than "OFF". ②When the display is clear without "BATT" showing, proceed to measurement. ③When the display blanks or "BATT" is indicated, replace the batteries according to section 8: battery replacement. NOTE ● It is possible that display is kept in blank while the function selector switch is set to a position other than "OFF".
6.Measurement 6−1 Current Measurement WARNING ●Do not make measurement on a circuit above 750V AC. This may cause shock hazard or damage to the instrument or equipment under test. ●Transformer jaw tips are designed not to short the circuit under test. If equipment under test has exposed conductive parts, however, extra precaution should be taken to minimize the possibility of shorting. ●Do not make measurement with the battery compartment cover removed from the instrument.
③Take the reading on the display. NOTE ● During current measurement, keep the transformer jaws fully closed. Otherwise, accurate measurement cannot be made. The maximum measurable conductor size is 54.5mm in diameter. ● When measuring a larger current, the transformer jaws may buzz. This is not a fault and does not affect the accuracy at all. 6−1−2 Peak Current Measurement ①Set the function selector switch to the "∼400A" or "∼2000A" position. ②Press the mode switch to select the peak mode.
6−2 Voltage Measurement WARNING ●Do not make measurement on a circuit above 750V AC or 1000V DC. This may cause shock hazard or damage to the instrument or equipment under test. ●Do not make measurement with the battery compartment cover removed. Black test lead Red test lead 6−2−1 DC Voltage Measurement ①Set the function selector switch to the " V" position. ②Slide the terminal cover to the left. Plug the red test lead into the V/Ω terminal and the black test lead into the COM terminal.
6−3 Resistance Measurement WARNING ●Before attempting to make measurement, make sure that the circuit under test is not live. The instrument is protected against a voltage up to 600V. ●Do not make measurement with the battery compartment cover removed. Black test lead Red test lead 6−3−1 Resistance Measurement (Normal Mode) ①Set the function selector switch to the "Ω/ " position. ②Slide the terminal cover to the left.
6−3−2 Continuity Check ①Set the function selector switch to the "Ω/ " position. ②Slide the terminal cover to the left. Plug the red test lead into the V/Ω terminal and the black test lead into the COM terminal. ③Press the mode switch to set the instrument to the continuity check mode. The measuring range is fixed to 400Ω and the " " symbol is shown on the display. ④Short the tip of the test leads and make sure that the display reads "0" and the buzzer beeps.
7.Notes on Functions 7−1 Data Hold This is a function used to freeze the measured value on the display. ①Press the data hold switch. The reading becomes frozen and the "H" symbol is shown on the display, indicating the instrument in the data hold mode. ②To exit the data hold mode, press the data hold switch again to release it. NOTE ● When the function selector switch is turned while the instrument is in the data hold mode, the data hold function remains activated.
②To enable the sleep function, turn the function selector switch back to "OFF", then to any other position. NOTE ● The instrument consumes small amount of current in the sleep mode. When the instrument is not in use, make sure to set the function selector switch to "OFF". 7−3 Recorder Output Only on the "∼400A" or "∼2000A" range, DC voltage proportional to the input current is output from the OUTPUT terminal. ①Set the function selector switch to the "∼400A" or "∼2000A" position.
[How to attach a pair of leads to the recorder output plug] Attach one end of the leads to the recorder output plug observing correct polarity, then connect the other end to the recorder. DANGER Never apply voltage to the output terminal. 7−4 Mode Switching Function On a AC current ("∼400A" or "∼2000A") or the resistance(Ω/ ) range, press the mode switch to cycle through the measurement modes.
8.Battery Replacement WARNING ●To avoid electric shock hazard, make sure to set the function selector switch to "OFF" and remove the test leads from the instrument before trying to replace batteries. CAUTION ●Do not mix new and old batteries. ●Make sure to install batteries in correct polarity as indicated in the battery compartment. NOTE ● If the instrument is powered on, but the display blanks or "BATT" is shown on the display, replace the batteries.
9.Optional Accessories ● Multi-Tran MODEL 8008 extends the capability of MODEL 2002PA or MODEL 2002R, allowing measurement up to 3000A or on a large bus-bar or conductor. ①Set the function selector switch to "∼400A". ②As shown in the figure below, clamp MODEL 2002PA or MODEL 2002R onto the pickup coil of MODEL 8008. ③Clamp MODEL 8008 onto the bus-bar or conductor under test. ④Take the reading on MODEL 2002PA or MODEL 2002R and multiply it by 10.