6−2 Voltage Measurement
WARNING
Do not make measurement on a circuit above 750V AC or 1000V
DC. This may cause shock hazard or damage to the instrument or
equipment under test.
Do not make measurement with the battery compartment cover
removed.
6−2−1 DC Voltage Measurement
Set the function selector switch to the " V" position.
Slide the terminal cover to the left. Plug the red test lead into the V/Ω
terminal and the black test lead into the COM terminal.
Connect the other end of the test leads to the circuit under test.
Take the reading on the display. When the red lead is the negative
potential, the "" sign is shown on the display.
6−2−2 AC Voltage Measurement
Set the function selector switch to the "V" position.
Slide the terminal cover to the left. Plug the red test lead into the V/Ω
terminal and the black test lead into the COM terminal.
Connect the other end of the test leads to the circuit under test. Take
the reading on the display.
NOTE
For high sensitivity, there are parts which do not indicate "".