Owner manual

―39―
6−3 Resistance Measurement
WARNING
Before attempting to make measurement, make sure that the
circuit under test is not live. The instrument is protected against a
voltage up to 600V.
Do not make measurement with the battery compartment cover
removed.
6−3−1 Resistance Measurement (Normal Mode)
Set the function selector switch to the "Ω/" position.
Slide the terminal cover to the left. Plug the red test lead into the V/Ω
terminal and the black test lead into the COM terminal.
Short the tip of the test leads and check whether the display reads "0".
Connect the tip of the test leads to the circuit under test. Take the
reading on the display.
NOTE
When shorting the tip of the test leads, the display may read a very
small resistance instead of "0." This is the resistance of the test
leads, not a fault.
If one of the test leads is open, the display reads "OL".
Black
test lead
Red test lead