6−3−2 Continuity Check
Set the function selector switch to the "Ω/ " position.
Slide the terminal cover to the left. Plug the red test lead into the V/Ω
terminal and the black test lead into the COM terminal.
Press the mode switch to set the instrument to the continuity check
mode. The measuring range is fixed to 400Ω and the " " symbol
is shown on the display.
Short the tip of the test leads and make sure that the display reads
"0" and the buzzer beeps.
Connect the tip of the test leads to the circuit under test. The
display reads the resistance and the buzzer beeps when the
reading is not more than about 50Ω.
NOTE
When shorting the tip of the test leads, the display may read a very
small resistance instead of "0". This is the resistance of the test
leads, not a fault.
If one of the test leads is open, the display reads "OL".