Owner manual

―17―
Connect the tip of the red and black test leads to the circuit under test
and take the reading on the display.
NOTE
When the voltage under test measures 3% of the full scale or less,
or the frequency of the voltage is low, "LoHz" is indicated on the
display.
6−5 Resistance Measurement
DANGER
Never use the instrument on an energized circuit.
Do not make measurement with the battery compartment cover
removed.
Keep your fingers and hands behind the barrier during measurement.
Set the function selector switch to the " Ω" position.
Slide the terminal cover to the left to disclose the V/Ω and COM
terminals. Plug the red test lead into the V/Ω terminal and the black
test lead into the COM terminal.
With the tip of the test leads shorted together, press the Zero
Adjust/Reset button to offset the resistance of the test leads.
Connect the tip of the test leads to the circuit under test and take the
reading on the display.