User Manual
―8―
Power Source :One 6 F22 or equivalent battery
Low Battery Warning :“B”symbol is shown on the display.
Current Consumption :Approx. 5mA max.
Standard(Safety) :IEC61010-1
IEC61010-2-032
CAT.Ⅲ,AC300V,Pollution degree 2
(EMC) :IEC61326
EN55022
Overload Protection :1500A AC max for 1 minute
Withstand Voltage :3700V AC for 1 minute between
electrical circuit and housing cases
Insulation Resistance :10MΩ or greater at 1000V between
electrical circuits and housing cases or
metal parts of jaws
Conductor Size :Approx. 68mm diameter max.
Dimensions/Weight :250(L)×130(W)×50(D)mm×. 570g
(including battery )
Accessories :6F22 battery
Carrying case
Instruction manual
Optional Accessories :Model 7073 two-way output cord
Peak Hold :Response time selectable from approx.
10ms and 100ms.
1/√2 of the peak current is displayed,
which means an RMS reading will be
obtained when current having
sinusoidal waveform in measured.
Peak hold reading varies by
approximately ±1% max of full scale
in the first minute at 23℃±10℃ and
relative humidity of 75% without
condensation.