●DC Output
*DC voltage is output in proportion to the display reading, which
reflects frequency selector or peak hold switch position200mV
DC for 2000 count reading.
200mA
2A
20A
200A
1000A
0∼200mA
0∼2A
0∼20A
0∼200A
0∼500A
501A∼1000A
Ranges
AC Input
Current
DC Output
Voltage
Accuracy (Frequency Range)
Freq. Selector
Switch at
WIDE position
Freq. Selector
Switch at 50/60Hz
Position
0∼200mV
±3%rdg
±3.5%rdg
±5%rdg
±7%rdg
±3.5%rdg
±4.0%rdg
±5.5%rdg
±7.5%rdg
0∼50mV
50∼100mV
Operating System Dual Integration
Display Field effect 3 1/2 digit liquid crystal
display with maximum count of 1999
Range selection Manual
Overrange Indication :“1is displayed on the highest digit
except for 1000A AC range.
Response Time Approx. 1 seconds
Sample Rate Approx. 3 times per second
Data Hold Available on all ranges
Location for use Indoor use, Altitude up to 2000m
Temperature & Humidity
for Specified Accuracy 23℃±10, relative humidity up to
85% without condensation
Storage Temperature
& Humidity :−1050, relative humidity up to
80% without condensation
Operating Temperature
& Humidity 040,relative humidity up to 85%
without condensation