Manual

IEC 60529 (IP54) Dust & drip proof
Operating System: Dual integration
Display: Liquid crystal display (maximum count: 1999),
Unit, Mark
Bar graph 30 points max. (20 points on
range)
Over range Indication: "OL" is shown on the display.
Sample Rate: Approx. 0.5 ~ 2.5 times per second
Operating Temperature 0 ~ +40
°C, relative humidity up to 85%
& Humidity:
Storage Temperature -20 ~ +60
°C, relative humidity up to 75%
& Humidity:
Insulation Resistance:
More than 50M at 1000V DC between electrical
circuit and housing case
Withstand Voltage: 3700V AC for one minute between electrical
circuit and housing case
Overload Protection
Insulation resistance ranges:
1000V Range 1200V (DC+AC p-p) for 10 seconds
500V Range 600V (DC+AC p-p) for 10 seconds
250V Range 300V (DC+AC p-p) for 10 seconds
Continuity ranges:
20/200/2000 280V (DC+AC p-p) for 10 seconds
Range (Protection by fusing)
7
Typical Number of Measurements.
(central tendency for supply voltage up to 8V)
Insulation Resistance Ranges:
Approx. 1000 times min. at load 0.5M
Continuity Ranges:
Approx. 700 times min. at load 1
Operating Errors (IEC 61557-2,-4)
Functions Range
Measuring range to keep Maximum percentage
operating error operating error
20M 0.50 ~ 19.99M
1000V 200M 1.0 ~ 199.9M
2000M 10 ~ 1000M
20M 0.50 ~ 19.99M
500V 200M 1.0 ~ 199.9M
2000M 10 ~ 100M
20M 0.25 ~ 19.99M ±30%
250V 200M 1.0 ~ 199.9M
2000M 10 ~ 100M
20 0.20 ~ 19.99
200 1.0 ~ 199.9
2000 10 ~ 1999
The influencing variations used for calculating the operating error are
denoted as follows;
Temperature: 0and 35
Supply voltage : 8V to 13.8V
Applicable Standards
IEC 61010-1 Over voltage CAT III 300V Pollution Degree 2
IEC 61010-2-31 Safety requirements for hand-held probe
assemblies
IEC 61557-1/2/4 Measuring equipment for low voltage
distribution systems
IEC 61326-1 EMC