Manual

3
KEW 3121A KEW 3122A KEW 3123A
DC Test Voltage 2500V 5000 V 5000V 10000V
Measuring Ranges
02000MΩ/
1000100000MΩ
automatic change
05000 MΩ/
2000 200000MΩ
(automatic change)
05GΩ/2200GΩ
(automatic change)
010GΩ/4400GΩ
(automatic change)
Accuracy
Insulation
Resistance
±5% of reading
10050000MΩ)
±10% of reading
or 0.5% of scale length
ranges other than
listed above
at
23 ±5
±5% of reading
200 100000MΩ)
±10% of reading
or 0.5% of scale length
ranges other than
listed above
at 23 ±5
±5% of reading
0.2100GΩ)
±10% of reading
or 0.5% of scale length
ranges other than
listed above
at 23 ±5
±5% of reading
0.4200GΩ)
±10% of reading
or 0.5% of scale length
ranges other than
listed above
at 23 ±5
±10% of reading
10050000
MΩ)
±20% of reading
or 1.0% of scale length
ranges other than
listed above
at 10℃∼+40
±10% of reading
200 100000MΩ)
±20% of reading
or 1.0% of scale length
ranges other than
listed above
at 10 ∼+40
±10% of reading
0.2100GΩ)
±20% of reading
or 1.0% of scale length
ranges other than
listed above
at 10℃∼+40
±10% of reading
0.4200GΩ)
±20% of reading
or 1.0% of
scale length
ranges other than
listed above
at 10℃∼+40
Output
Voltage
2500V ±5%
10050000MΩ)
5000V ±5%
200 100000MΩ)
5000V ±5%
0.2~100GΩ)
10000V ±5%
0.4~200GΩ)
Operating
Temperature & Humidity
10 ∼+40 at 85% max. relative humidity
Storage
Temperature & Humidity
20 ∼+60 at 90% max. relative humidity
Insulation Resistance 1000MΩ max./ 1000V between electrical circuit
& housing case
Withstand Voltage 5000V AC for one minute between electrical circuit & housing case
Dimensions 200L×140W×80Dmm
Weight Approx. 1kgincluding batteries
Power Source 8 pcs of R6P (AA) battery or equivalent
Accessories
Hard Carrying Case:M-9158 Test Leads
(Line Probe:M-7165A,
Earth
Cord:M-7224A, Guard Cord:M-7225A)
Hard Carrying Case:M-9158 Batteries, Test Leads (Line
Probe:M-7165A, Earth Cord:M-7224A, Guard Cord:M-
7225A) Pickel type prod:M-8019
Optional adaptor model 8324 is available for connection to recorder