Instruction Manual

̶
8
̶
●Overload protection:
Insulation resistance range: AC1200V/10sec.
Voltage range: AC720V/10sec.
●Withstand voltage: AC5320V(50/60Hz)/5sec.
(Between electrical circuit and enclosure)
●Insulation resistance: 1000MΩ or more/DC1000V
(Between electrical circuit and enclosure)
●Dimension: 205(L)×152(W)×94(D)mm
●Weight: approx. 1.8kg (battery included)
●Power source: DC12V: Alkaline battery size C(LR14)x 8pcs
Current consumption (representative values at 12V of supply voltage)
Range 500V 1000V 2500V 5000V V
AC/DC
Output at
short-circuit
For 10 sec after a test is started :
850mA, after that : 150mA
110mA
* at voltage
measurement
When rated
measuring current
is outputted
450mA
/0.5MΩ
500mA
/1MΩ
600mA
/2.5MΩ
850mA
/5MΩ
Output at
open circuit
45mA 50mA 70mA 110mA
On stand-by 25mA 110mA
When backlight
is on
Increased by 30mA
Note) Current values in above table are all approximate values.
●Operating time: approx. 10 hours continuous
-
under a load of 100MΩ at 5000V Insulation
resistance range
●Accessories: Line probe: MODEL7165A
Earth cord: MODEL7224A
Guard cord: MODEL7225A
Alkaline size C battery (LR14) x 8pcs
Instruction manual
Hard Case MODEL9159
Pickel Type Prod: MODEL8019
Straight Type Prod: MODEL8254
●Optional accessories: Adaptor for recorder MODEL8302
Line probe with alligator clip MODEL7168