Owner manual

12 KEW3128
3. Specification KEW3128
Volt Meter
Range Voltage Test
Measuring Range
DC Voltage AC Voltage
±30~±600V 30~600V(50/60Hz)
Accuracy ±2%rdg±3dgt
Frequency
Range Voltage Test
Measuring Range 45.0 ~ 65.0Hz
Accuracy ±0.2Hz
Ammeter
Measuring Range
0.0nA ~ 2.40mA
(determined by
resistance and voltage
values)
Depending on the
effective range of
insulation resistances
Capacity Meter
Range 500V ~ 5000V Range 10000V / 12000V Range
Accuracy
5.0nF ~ 50.0μF 5.0nF ~ 1.0μF
±5%rdg±5dgt
* Measurements of capacitance of 0.5μ or more shouldnt be repeated at
10000V/ 12000V Ranges in a short time. (Rough guide: 5 times/ hour)
* A message Noise Error may be displayed on the LCD and a measurement
may be ceased at 10000/12000V Ranges. In this case, select a lower Voltage
Range and test again.
Calculated Value
PI,DAR,DD
Measurement Mode PI DAR DD
Display Range 0.00 ~ 999 0.00 ~ 999 0.00 ~ 999
Computational error ±2dgt ±2dgt ±2dgt
Electromagnetic compatibility (IEC61000-4-3)
Radio-frequency electromagnetic field
= 10V/m : 20times of the specified accuracy