Owner manual

KEW3128 47
KEW3128 6. Measurement
6. 2 Insulation Diagnosis Tests
   This instrument can measure and perform following items as a part of
Insulation resistance Test.
  - Insulation Resistance (IR)
  - Polarization Index (PI)
  - Dielectric Absorption Ratio (DAR)
  - Dielectric Discharge (DD) *Auto-testing
  - Step Voltage Test (SV)
Measurement Mode Function
Insulation
Resistance IR
Performs normal insulation resistance
measurements (consistent measurements)
Polarization IndexPI
Measures resistances twice and calculates
polarization index automatically.
default value: 1 min, 10 min
Dielectric Absorption
Ratio DAR
Measures resistances twice and calculates
dielectric absorption ratio automatically.
default value: 15 sec, 1 min
Dielectric Discharge
DD
Calculates dielectric discharge based on the
measured capacitance of the measured object and
residual current values after testing.
Step Voltage TestSV
Increases the set voltage by 20% every time
when pre-set time comes.