Owner's manual
̶
4
̶
3. Specification
Operating system Earth resistance function: Constant voltage injection,
Current detection,
(Frequency: Approx.
2400
Hz)
Dual Integration
AC current function:Successive Approximation(True-RMS)
Display Liquid crystal display with a maximum count of
2099
Over-range indication "OL" is displayed when input exceeds the upper limit
of the measuring range
Response time Earth resistance function :Approx.
7
seconds
AC current function :Approx.
2
seconds
Sample rate Approx. once per second
Location for use Altitude
2000
m or less, Indoor/ outdoor use
IP protection degree IP
40
Temperature & humidity range
23
℃ ±
5
℃
/Relative humidity
85
% or less
(guaranteed accuracy)
(
no condensation
)
Operating temperature & -
10
℃
to
40
℃
/Relative humidity
85
% or less
humidity range (no condensation)
Storage temperature & -
20
℃
to
60
℃
/Relative humidity
85
% or less
humidity range
(
without batteries, no condensation
)
Power source DC
6
V: R
6
P (size AA manganese battery) x
4
pcs, or
LR
6
(size AA alkaline battery) x
4
pcs
Current consumption MODEL
4200
: Approx.
50
mA (max.
100
mA)
KEW
4202
: Approx.
90
mA (max.
140
mA)
Measurement time MODEL
4200
: Approx.
12
hours (when using R
6
P), or
24
hours (when using LR
6
)
KEW
4202
: Approx.
5
hours (when using R
6
P), or
21
hours (when using LR
6
)
Measuring range and accuracy
Function Range Resolution
Measuring range
Accuracy
Earth resistance
(Auto-ranging)
20
Ω
0
.
01
Ω
0
.
00
-
20
.
00
Ω ±
1
%
±
5
dgt
200
Ω
0
.
1
Ω
20
.
0
‒
200
.
0
Ω ±
3
%
±
5
dgt
600
Ω
1
Ω
200
-
399
Ω ±
6
%
±
5
dgt
400
-
599
Ω ±
10
%
±
10
dgt
1200
Ω
10
Ω
600
-
1260
Ω
-
AC current (ACA)
(
50
Hz/
60
Hz)
(Auto-ranging)
100
mA
0
.
1
mA
0
.
0
-
100
.
0
mA
±
2
%
±
5
dgt
(sine wave)
1000
mA
1
mA
100
-
1000
mA
10
A
0
.
01
A
1
.
00
-
10
.
00
A
30
A
0
.
1
A
10
.
0
-
30
.
0
A
* Crest factor
≦
2
.
5
Accuracy at sine wave +
1
% (
50
Hz/
60
Hz, peak value shall not exceed
60
A)
* In the following cases, zero will be displayed on the LCD.
- At
20
Ω
range of Earth resistance function:
0
.
04
Ω
or less
- At
100
mA range of ACA function:
0
.
4
mA or less
* A range shifts to upper range when the input exceeds
105
% of the selected range, and shifts
to the lower range when the input falls under
80
% of the range.