8. Now Setting 1 is completed; "End" is displayed on
the LCD. Press the button to return to the
screen on which "SEt.1" is displayed.
9.
Press the
button to get the instrument ready
for a measurement.
Max number of recorded data
* Setting is available only on PC.
Using only 1 channel
345 data
Setting item
Max number of recorded data
* Setting is available only on PC.
Using only 1 channel
4000 data
Setting item
Capture recording mode: (Setting on PC)
Power Quality Analysis [only 5020] (Setting on PC)
Item Range for Setting Default
CH1 Detection Level 0 ~ 1000
15
One-time/ Endless One-time/ Endless Endless
Item
Range for Setting
Default
Reference Voltage (System Supply Voltage) 100V
500V 100V
Swell Detection (Ratio against Reference Voltage)
100%
200% 110% (110V)
Dip Detection (Ratio against Reference Voltage)
0%
100% 90% (90V)
Short Interruption Detection
(Ratio against Reference Voltage)
0%
100% 10% (10V)
Hysteresis (Ratio against Reference Voltage)
0%
10% 1% (1V)
One-time/ Endless
One-time/ Endless
Endless