Operating system : Successive Approximation
(CH1 single synchronized sampling)
󱛠
Rated max. working voltage : AC9.9Vrms, 14V peak value
󱛠
Number of input channel : 3ch
󱛠
Measuring method : True RMS
󱛠
RMS measuring interval : approx. 100ms.
󱛠
Sampling interval
Normal/Trigger mode : approx. 1.65ms/CH
Capture mode : approx. 0.55ms (waveform: at every 1.1ms)
P.Q.A mode : approx. 0.55ms
󱛠
Display : Liquid crystal display
󱛠
Low battery warning : Battery mark display (in 4 levels)
󱛠
Over-range indication : "OL" mark is displayed when exceeding the
measuring range.
󱛠
Auto power off : Power-off function operates automatically after
a switch remains for 3min. (when recording is
stopped)
󱛠
Location for use : Indoor use, Altitude up to 2000m
󱛠
Temperature & humidity range(guaranteed accuracy):
23
±
5
/ Relative humidity 85% or less (no
condensation)
󱛠
Operating temperature & humidity range:
-10
~50
/ Relative humidity 85% or less (no
condensation)
󱛠
Storage temperature & humidity range:
-20
~60
/ Relative humidity 85% or less (no
condensation)
󱛠
Battery : DC6V: Alkaline battery(LR6) x 4pcs
External supply DC9V (Special AC Adapter)
󱛠
Current consumption : approx. 10mA
󱛠
Possible measurement time
: Approx.10days (with alkaline LR6 batteries)
󱛠
Applicable standards : IEC 61010-1:2001 CATI II 300V Po llu tio n
degree2
IEC 61326 (EMC standard)
󱛠
Overload protection : 120%MAX/ 10sec of rated current/ voltage of
each Sensor
󱛠
Withstand voltage : AC3540V (RMS 50/60Hz)/ for 5sec.
󱛠
Insulation resistance : 50Mohm or more / 1000V