Instruction Manual User Manual
6.11 KEW6305
File No.
6.4 Saving data (instantaneous values)
Pressing the SAVE key on W range during a measurement saves all the measured parameters at the
instant of saving. This is a manual single step operation.
Data can be saved to either below two locations:
* SD card : Max. 511 files can be saved.
* Internal memory : Max. 4 files can be saved.
Data is saved to a SD card automatically when a SD card has been inserted. If a SD card has not been
inserted, data is automatically saved to internal memory.
6.4.1 Saving Procedure
1 Press the SAVE key whilst on the W range .
2 The File number screen appears and the instantaneous measured data is saved.
(A file number is assigned automatically.)
3 On the measurement screen, it can be seen that a file is open.
〈 SD card〉
lights up
〈Internal memory〉
lights up
File number
screen
Measurement
screen
4 Subsequent measured data can be saved by pressing the SAVE key with a file already opened.
5 Closing a file. On completion of the data gathering, the file has to be closed. Set the Function switch to
any range other than “OFF” and W .(eg. WIRING CHECK )