Quick Manual Instruction Manual

22
KEW6305
KEW6305
How to start/ stop measurement
Start measure-
ment
LED
LED blinks
Stand-by
mode
8-2. How to start/ stop measurement
To start/ stop measurement manually To start/ stop measurement at the preset date&time
Hold down the
START
/STOP
Key for
over 2 sec.
DEMAND
Range
Configure the settings for Setting
14/15, and then press the
START
/STOP
Key.
File no. screen is displayed for about 2 sec.
- Manual operation : Pressing the
START
/STOP
Key starts measurement anytime.
- Automatic operation : KEW6305 enters into stand-by mode and wait until the
preset start time comes.
The Status indicator LED blinks when the instrument is in stand-by mode and it
stays on when measurement starts.
* Display is updated every 1 sec.
* Data will be saved at the interval selected at Setting 13.
* Buzzer sounds when a predicted value exceeds a target value set
at Setting 15 in the pre-set period set at Setting 17.
* While performing measurement, select the W Range to view
instantaneous values, Wh Range to view integration values and
SET UP Range to check the details of settings.
Manual operation
Hold down the
START
/STOP
Key for over
2 sec.
Automatic operation
Measurement stops automatically
when the preset date& time comes.
The status indicator LED goes off when measurement stops (or file is closed).
* When the destination to save data has been set to the internal memory,
symbol is displayed
on the screen instead of symbol.
Integration values remain displayed on the LCD when measurement ends.
Press the
ESC
Key at least 2 sec to select dEL and clear the display if the displayed values will not
be used in further measurements.
LED lights up
Measuring
Stop measurement