User Manual

KEW6310 8.1.1 Demand Measurement – Display Screen
KEW6310 8.2
Measurement screen
Displaye d
para me te r s
Detail s
Remaining time
(time left)
Demand in terval is coun ted down.
Target value Should be set for each measurement.
Predicted value
Predicted demand value (average power) when preset demand interval elapses
under present load .
(Present value) x (Preset interval)
(Elapsed time)
* Integration and calculations are done as time elapses.
Present value
Demand value (average power) within a demand interval.
“WP+ x 1 hour”
Interval
* Integration and calculations are done as time elapses.
Max demand
Max demand recorded in a measuring period is displayed. Displayed value will be
refreshed if any highe r de mand is dete cte d.
Remaining time (time left)/ Target value/ Predicted value / Present value
Measured max de man d
with time and date
informatio n