User Manual

KEW6310 11.2.2 Power quality – Measuring Procedure (Swell/ Dip/ Int measurement)
KEW6310 11.6
Recording at every interval
.* Function available with ver2.00 or later .
Interval
Inst values* are recorded at every interval.
Ave,max and min* values within each interval are
recorded.
Inst value : Avg of 100 data (@50Hz) obtained in the preset inst interval of 1 sec (RMS)
A vg value : Avg of rms values obatained in the preset inst interval
Max value : Max rms values obtained in the preset inst interval
Min value : Min rms values obtained in the preset inst interval