User Manual
11.4.2 Power quality – Measuring procedure (Inrush current measurement) KEW6310
11.21 KEW6310
Timing of data recording
<Recording at event occurrence>
<Recording at every interval>
* Function available with ver2.00 or later .
Start
End
100%
(Reference)
current
110%
( Threshold
vallue)
109%
(Thresholc value110%-Hysteresis1%)
100
data pts
100
data pts
Tri
gg
e
r
Trigge
r
100
data pts
100
data pts
201 data pts
recorded
Duration
RMS(max)
201 data pts
recorded
Interval
Ave,max and min values within
each inte rval are record in
g
.
※Inst value : Avg of 100 data (@ 50Hz) o btained in th e preset inst
interval of 1sec(RMS)
A vg value : Avg of rms values obtained in the preset inst interval
Max value : Max rms values obtained in the preset inst interval
Min value : Min rms values obtained in the preset inst interval.
Inst values recorded at