User Manual
KEW6310 12.4 Backup memory
KEW6310 12.10
Dat a processing in Backup memory
One CF card is placed/ removed during saving data;
Saving
① A file is created in the C F card , w hen C F card i s se lecte d as a de sti natio n fo r sa ving da t a , and
measurement data is saved to the CF card.
②
A backu p file is created in the inte rnal me mo ry when a CF card is removed at sa ving data.
Further data is saved to the internal memory.
③
When inserting the CF card again during a data saving, further data will be saved to the last
available space in CF card. (behind the ①& ②).
KEW6310
Internal memory
CF card
01-CF001.csv
②
①
KEW6310
CF card
Internal memory
01BCF001.csv
③
CF card
01-CF001.csv
②
①
KEW6310
Internal memory
01BCF001.csv
③
CF card
01-CF001.csv
①
②
④
⑤