User Manual

KEW6310 14.1 Input/ Output terminals
KEW6310 14.2
[Input terminal]
Capable of measuring and recording DC voltage signals.
Number of Ch: 2ch
Input resist ance : approx 225 k
CAUTION
Roots of the L terminals for each Ch are integrated. Never connect inputs with various grand levels to
the terminal at the same time.
[Output terminal]
Capable o f genera ti ng ou tpu t s w hen eve nts occur during mea sure ments below.
Measurement menu Conditions for Output : Lo Remarks
Demand
(Predicted value) > (Target value) ------
Harmonic
exceeding preset allowable
range
Lo outpu t; when an alowable
range i s exceeded at any Ch.
Swell/ Dip/ Int/ Tr ansient/
Inrush
new event i s added and
displayed on the LCD
Lo is kept for 1 sec, Hi is
restored
Unbalance rate
exceeding pre se t th reshol d ------
Output format : Open collector output
Max input : 30V, 50mA, 200mW
Output voltage : Hi – 4 ~ 5V
Lo – 0 ~ 1V
+5V
H
L
OUTPUT