14.2 Getting power from measured lines
When there is difficulties in getting power from an outlet, KEW6310 operates with powers on the
measured line by using V oltage test leads with Power supply adapter MODEL8312.
Connect the Adapter according to following procedure.
1 Confirm that the Adapter is off.
2 Connect the Plug of the Adapter to VN and V1
terminals on KEW6310 /6300 and Power Plug to
the Power co nnecto r resp ecti vely.
3 Connect the Voltage test leads to VN and V1
terminals of the Adapter.
4 Connect the Alligator clips of the Voltage test
leads to the circuit under test.
5 Power on the Adapter.
6 Power on KEW6310/ 6300.
* Reversed procedure is applied to remove the
Adapter from KEW6310/ 6300.
Fuse rating : AC500mA/ 600V ,
Fast acting, Φ6.3 x 32mm
For further details, refer to the Instruction manual for MODEL 8312.