User Manual

16.9 Specification of Clamp sensor KEW6310
16.13 KEW6310
MODEL8141 MODEL8142 MODEL8143
Rated current AC1000mA
Output volt age
AC0 ~ 100mV
(AC100mV/ AC1000mA)
Measuring range AC0 ~ 1000mA
Accuracy
(sine input)
±1.0%rdg±0.1mV (50/ 60Hz)
±2.0%rdg±0.1mV (40Hz ~ 1kHz)
Phase
characteristics
------------- -- --- -
Temp. & humidity
range (guaranteed
accuracy)
23±5°C, relative humidity 85% or less (no condensation)
Operating temp.
range
0 ~ 50°C, relative humidity 85% or less (no condensation)
Storage temp. rang e
-20 ~ 60°C, relative humidity 85% or less (no condensation)
Allowable input AC100Arms (50/60Hz) AC200Arms (50/60Hz) AC500Arms (50/60Hz)
Output impedance approx 180 approx 200 approx 120
Location for use indoor use, altitude 2000m or less
Applicable standard IEC 61010-1, IEC 61010-2-032
Measurement CAT.III (300V) Pollution degree 2
IEC61326 (EMC standard)
Withstand voltage AC3540V / 5 sec
between Jaws - enclosure
between enclosure – output terminal
between Jaws – output terminal
Insulation
resistance
50M or more/ 1000V
between Jaws - enclosure
between enclosure – output terminal
between Jaws – output terminal
Max conductor size Φ24mm Φ40mm Φ68mm
Dimension
100(L) × 60( W) × 26 (D)m m
(protrusions are not included)
128(L) × 81( W) × 36 (D)m m
(protrusions are not included)
186(L) × 129( W) × 53( D)m m
(protrusions are not included)
Cable len gth approx 2m
Output terminal MINI DIN 6PIN
Weig ht approx 15 0g appr ox 24 0g approx 49 0g
Accessory Instruction manual,
Carrying case
Option 7146 (Φ4 Banana plug),
7185 (Extension lead)