4.2.1 Basic Setting KEW6310
4.11 KEW6310
<< Auto Setting >>
Model name of the Clamp sensor connected to the Current Terminal of the instrument is detected auto-
matically at Auto setting mode. Setting for [Wiring] should be don e to ad vance Auto setting.
1 Confirm that settings for [Wiring] are made, and then press the F2 Key.
The max measurable values on Clamp sensor are reflected in setting for [A Range].
[CT ratio] is automatically set to 1.00.
For [Filter], bars are displ ayed when the detected senso rs are MODEL8 12X series and OFF is
displayed when the sen sors ar e MODEL814X se ries .
Setting will be changed if new sensors are detected at powering on the instrument.
Pressing the F2 Key initiates
auto setting for Clamp sensor.
Connected Clamp sensors are automatically
detected, and se tting s fo r [A Range], [CT ratio]
and [Filter ] are ma de auto ma tica lly.