KEW6310 4.2.1 Basic Setting
KEW6310 4.12
The instrument detects and checks the connected Clamp sensors and the selected wiring configuration, and
displays following messages when improper Clamp sensors are connected.
< Improper Clamp sensor is detected >
Recheck and connect p roper Cla mp sen sors.
< No sensor is detected >
Check the Clamp sensor connected to the Current input terminal corresponding to the Ch number
displayed with question ma rk.
When starting measurement with the question mark displayed at the [Clamp] box, previous setting
is applied auto ma tica lly.
Model names of the connected
Clamp sensors are display ed. Bars
are displayed at A range, CT ratio
and Filter boxes.
Question mark “?” is displayed at Model
name and Curren t ra nge box e s. C T ra tio is
automa ti ca lly s et to 1 .00 . Bars are displ ayed
at Filter box.
Question mark “?” is displayed at Model
name box.
CT ratio is automatically set to 1.00.
Bars are displayed at Filter box.