User Manual

4.2.2 Measurement setting (WAVE Range) KEW6310
4.29 KEW6310
WAVE Range Setting
Press the F2 Key at each Measurement setting screen, and move to the screen for W AVE Range Setting.
Setting for interval
* Default value (or after system reset) : 30 min
* Setting procedure is same to that for interval Setting for W/ Wh/ DEMAND.Refer to the procedure described in
the preceding pages.
Interval listed on the drop down list depends on the number of save items with “ON” setting. Alter
the number of save items with “ON” setting when desirable interval isn’t listed on the drop down list.
Interval Number of “ON”
1 sec 1
2 sec 2 or less
5 sec or more 5 or less
Setting for saving Waveform data
Paramete rs with “ON” se tting w ill be sa ved .
V
A
ONOFF
* Default value (or after system reset) : ON (all items)
1 Press the Cursor Keys and sele ct a p aramete r to be ch anged , and then pre ss the
ENTER Key.
2 Press the Cursor Keys and select “ON” or “OFF”, and then press the ENTER Key.
Dro
p
down list a
pp
ears.
Setting of the selected item is displayed.