KEW6310 4.2.2 Measurement Setting (QUALITY-Swell, Dip, Int Measurement)
4.39 KEW6310
Setting for Transient
Volt age Range 70~150V 151~300V 301~600V 601~1000V
T ran sien t
(on 1V basis)
50~310Vpeak 90~630Vpeak 170~1270Vpeak 340~2000Vpeak
* Default value (or after system reset) : 210V
* Vrms value (Vpeak divided by √2 ) is auto ma tica lly ca lcul ate d w hen Vpe a k is se t.
1 Press the Cursor Keys and select [Transient], and then press the ENTER Key.
2 Press the Cursor Keys and alter the values, and press the ENTER Key
to fix it.
Box with ▲▼ mark appears
at the rightmost digit.
Selected threshold value is
displayed. Value displayed in
parenthesis is a Threshold value
divided by √2.