User Manual

KEW6310 4.2.2 Measurement Setting (QUALITY-Inrush Current Measurement)
KEW6310 4.52
1 Press the Cursor Keys and select [A_Referene], and then press the ENTER Key.
2 Press the Cursor Keys and alter the values, and press the ENTER Key to fix it .
Setting for Threshold
100 ~ 200% (can be se t by 1%)
* Default value (or after system reset) : 110%
* Setting procedure is same to that for Threshold Setting for Swell, Dip, int measurement.
Refer to the procedure described in the preceding pages.
Setting for hysteresis
1 ~ 10% (can be set by 1%)
* Default value (or after system reset) : 5%
* Setting procedure is same to that for Hysteresis Setting for Swell, Dip, Int measurement.
Refer to the procedure described in the preceding pages.
Setting for trigger point
Past: 0 ~ 200 (can be set by 1) Next : 200 ~ 0 (can be set by 1)
* Default value (or after system reset) : 100
* Trigger to st art and stop recordin g , when a preset th reshold exceed ed, w ill be decided based
on the number of recorded data.
* Setting procedure is same to that for Trigger Point Setting for Swell, Dip, Int measurement.
Refer to the procedure described in the preceding pages.
Box with ▲▼ mark appears
at the rightmost digit.
Selected reference current
is displa
y
ed.