Owner manual

Chapter 2: Prerequisites
Product Service 1-800-522-7658
7
Location Requirements
The Freeze Dryer should be located in an area that provides an unobstructed flow
of air around the cabinet. This air cools the refrigeration system. The
refrigeration system draws air in through the front and exhausts it through the
rear. A minimum of 3" must be allowed between the rear and both sides of the
Freeze Dryer and adjacent wall surfaces. Restriction of airflow during operation
could adversely affect performance.
Refer to Appendix B: Stoppering Tray Dryer Dimensions for dimensional
drawings of the Stoppering Tray Dryer.
Vacuum Pump Requirements
A vacuum pump must be provided by the user. A vacuum pump with a
displacement of 144 liters per minute and 0.002 mBar ultimate pressure is
adequate for most samples. The inlet fitting on the vacuum pump must be
suitable for 3/4" ID vacuum hose, which is provided with the FreeZone Freeze
Dry System. It is recommended that the vacuum pump is equipped with an
exhaust filter to minimize oil mist exhausting from the vacuum pump. The
operating vacuum level may be set on the Freeze Dry System. The higher the
pressure is set, the more likely it is that oil mist will be exhausted.
Vacuum pumps used with 115V models should be equipped with an 115V, 15
Amp NEMA 5-15P plug. Vacuum pumps used with 230V models should be
equipped with a reverse IEC plug. This plug is included with 230V models. This
will allow the vacuum pump to be plugged into the receptacle on the back panel
of the Freeze Dry System. Refer to the User’s Manual for the FreeZone Freeze
Dry System for vacuum pumps available from Labconco.