Chapter 4: Using Your Stoppering Tray Dryer
Product Service 1-800-522-7658
18
Hold for 5 seconds.
Press ENTER.
At any time the DISPLAY button can be pressed to select a different
display screen.
Manual Control Operation
(1) Turn the Power switch ON. The display will become active.
(2) Push the DISPLAY button until the Manual LED lights. This will
cause the manual display screen to be shown with the previously
entered set point temperature.
The temperature units are displayed in the units (°F or °C) selected in
the Set Up.
The temperature may be set anywhere from +40°C to 40°C in 1°C
increments.
(3) If the set point needs to be changed, press the INCREASE or
DECREASE button until the desired set point temperature is
displayed.
(4) Press ENTER. If ENTER is not pressed within 10 seconds, the set
point will revert to the previously entered set point temperature.
(5) Press MODE button until the Manual LED lights.
(6) Press RUN/STOP button. The amber LED by the button will light and
the refrigeration system and/or heater will operate.
(7) To change the set point temperature in the middle of a run, repeat steps
2, 3 and 4. The system will adjust to the new temperature set point.
(8) To monitor system operation, press DISPLAY until the Monitor LED
lights. The display will simultaneously show system temperature,
vacuum and three sample temperatures if sensors were plugged into
their jacks at the front of the chamber and inserted into sample vials.
If sensors are not plugged into their jacks, the probe temperature
display will show “- - -”.
SET POINT TEMP 30°C