Owner manual

Chapter 7: Troubleshooting
Product Service 1-800-522-7658
41
Shelf and Stoppering System Removal
(1) The shelf and stoppering system may be removed from the chamber as an
assembly for repairs if needed.
(2) Remove the acrylic door by taking out the two hex socket head screws
located on the right hand end of the cross bar.
(3) Remove the door gasket. Carefully pull loose all of the plastic fasteners
(located around the periphery of the chamber) from the chamber. Leave
the spacer blocks behind the gasket attached to the gasket.
(4) Place a support in front of the chamber and slide the shelf and stoppering
system out approximately 2 inches.
(5) Loosen the clamp securing the stoppering diaphragm tube (located in the
lower center at the back of the chamber) and remove the tube. Slide the
assembly out of the chamber and onto the support.
(6) The assembly is now ready for repairs. Whenever possible, do not break
the fluid system.
(7) Reverse the removal procedures for installation of the shelf and stoppering
system. Make sure the fluid inlet and outlet tubes are routed on top of the
assembly.