42
Filling the Fluid System
1. Fluid Circulation Valve
2. Fluid Reservoir and Cap
3. Electrical Junction Block
4. Flexible Tubing
(1) Make sure the Power Switch is in the OFF position.
(2) Remove the left cover.
(3) Remove the insulation and both caps from the fluid circulation valve,
which is located directly behind the fluid reservoir (Item 1).
(4) Attach a flexible tube to the exposed port (service port). The port is 3/8" x
45° male flare (Item 4).
(5) Remove the fluid reservoir cover (Item 2) and place the opposite end of
the tubing into the fluid reservoir.
(6) Fill the reservoir slowly allowing the fluid to drain through the system.
When the system appears full, prepare to start the circulation pump.