Lock Pin 
Stainless Ring 
•  6L, 12L and 18L Model Chambers 
o  Place the flat rubber gasket over the inlet port. 
o  Place the chamber over the gasket. 
o  Reaching into the chamber, set the stainless 
steel ring over the port opening. Slide the lock pin 
through the holes in the port and rotate over 
center (stainless steel chamber only). 
•  6L, 12L and 18L Manifolds- see instructions 
supplied with Manifolds. 
T
T
e
e
s
s
t
t
i
i
n
n
g
g
o
o
r
r
S
S
t
t
a
a
r
r
t
t
U
U
p
p
Step 6. Testing the System 
•  Connect the power cord into the receptacle on the back of the Freeze Dryer and insert 
the other end into a suitable wall receptacle. 
•  Turn the power switch ON. 
• Vacuum check 
o  Make sure the lid is on the collector and either a manifold 
or chamber is in place. 
o  Close all manifold valves. 
o  Set the vacuum level for .002 mBar (except 1L models 
without vacuum control). 
o  Press the Vacuum button. 
o  Vacuum should reach .133 mBar or less within 15 
minutes. 
•  Refrigeration (leave the vacuum pump running) 
o Press the Refrigeration button. 
o Collector temperature should reach operating temperature 
(-50°C or -84°C depending on the model) within 40 
minutes. 
•  If the Freeze Dryer fails to perform as expected, consult the User’s Manual for 
troubleshooting information or call Labconco. 