User Manual
Labconco Operating Instructions 5242600, Rev. B 4/27/07, ECO E275  13 of 20 
Labconco Glove Box Controller Specifications 
Glove Box 
Pressure Control 
Operating Range  Default Settings 
Pressure Control Limits  -5 inches of water column to 
+5 inches of water column; 
[+/- 12.6 mBar] 
+/- 5 inches of water column; 
[+/- 12.6 mBar]. 
Manual Control Available  Purge (Gas out); 
Fill (Gas in) 
Audio Alarm level  Whenever pressure 
outside limits 
Glove Box Purge/Fill 
Purge/Fill Pressure Limit  +/- 5 inches of water column; 
[+/- 12.6 mBar]. 
+/- 5 inches of water column; 
[+/- 12.6 mBar]. 
Manual Control Available  Purge (Gas out); 
Fill (Gas in) 
Number of Purge/Fill 
Cycles 
 0 to 199  1 
Audio Alarm level  Whenever pressure 
outside limits 
Transfer Chamber 
Purge/Fill 
Purge Pressure Limit  -29.5 inches of Mercury/Hg 
[-999 mBars] 
-29.5 inches of Mercury/Hg 
[-999 mBars] 
Fill Pressure Limit  0 inches of Mercury/Hg 
[0 mBars](atmosphere) 
0 inches of Mercury/Hg 
[0 mBars](atmosphere) 
Number of Purge/Fill 
Cycles 
0 to 199  1 
Audio Alarm level  -29.5 inches of Mercury/Hg 
[-999 mBars] 
Whenever pressure/vacuum 
outside limits 