Labconco Operating Instructions 5242600, Rev. B 4/27/07, ECO E275  17 of 20 
Benchtop Installation: With the Pressure Controller temporary located at the benchtop location 
proceed as follows. 
1.  Measure the distance between the solenoid valves of the Pressure Controller and the 
valves on the glove box. Cut the 3/16 ID tubing (the larger of the tubing in your kit) 
to the measured lengths. 
2.  The solenoid valves are embossed with IN or OUT port on the body of the valve, to 
indicate the direction of flow. Install the cut lengths of the tubing to the correct port 
of the solenoid valve as shown in Detail C. 
3.  Secure the tubes to the underside of the mounting bracket using the āCā clips 
provided and place at the approximate locations shown in Detail C. 
4.  Route the tubing to the Glove Box and Transfer Chamber as shown in Detail C. Note: 
Tube routing maybe dependent upon your application. Always insure that the 
tubing is routed to the correct solenoid valve on the Pressure Controller as 
stated on the back of the unit. 
5.  See Detail D for proper tube installation to the valves on the Glove Box and Transfer 
Chamber. 
Sensor Integration: 
The Pressure Controller must be integrated to the glove box. This connection 
allows your Pressure Controller to monitor the condition of the glove box 
chamber and transfer chamber. 
Installation 
1.  Your kit comes with extra length of polyvinyl tubing so you can integrate the 
Pressure Controller with the glove box. This integration requires you to splice into the 
existing tubing of your Glove Box, install a bulkhead fitting and new tubing. 
TUBE INSERT
TUBE FERRULE
BACK FERRULE
NUT
TUBING 