User Manual
Labconco Operating Instructions 5242600, Rev. B 4/27/07, ECO E275  5 of 20 
Main Modes of Operation 
The Labconco Glove Box Pressure Controller display has four main screens that provide access 
to the 4 modes of operation listed below. The operation modes are covered in the next sections of 
this document. 
1.) Glove Box Pressure Control Mode 
2.) Glove Box Purge and Fill Mode 
3.) Transfer Chamber Purge and Fill Mode 
4.) Setup Mode 
Press the Mode key at the upper-left hand corner of the display screen to switch between each of 
these four operation modes. Please be aware that if one of the glove box modes is 
running/started, other glove box operation modes cannot be used. For example, if the Glove Box 
Pressure Control mode has started (using the start key), the menu page for the Glove Box 
Purge/Fill mode cannot be used – this prevents operational control conflicts from occurring at the 
same time. 
Manual 
Purge 
Manual 
Fill 
Start/Stop 
Setup 
Increase 
Upper Limit 
Decrease 
Upper Limit 
Mode Key 
Mute/Alarm 
Increase 
Lower Limit 
Decrease 
Lower Limit 