Labconco Operating Instructions 5242600, Rev. B 4/27/07, ECO E275  7 of 20 
Glove Box Purge/Fill – Operation Mode 2 
The Glove Box Purge/Fill mode uses the same solenoid valves on the glove box that are used in 
the Pressure Control Mode. The Pressure Control mode will be temporarily suspended while a 
“Purge/Fill” command is being executed. 
Note that this display screen looks similar to the Pressure Control mode described previously. 
However, the “SETUP” key has changed to a “Cycles” key. Up to 199 cycles can be set for 
automatic “Purge/Fill” of the glove box main chamber by holding the “Cycles” key. To decrease 
the number cycles, press the “Cycles” key to 199 cycles and go around again until a lower cycle 
quantity appears. Use this display screen to also set the low pressure level for purging and the 
high pressure level for filling by adjusting the upper and lower limit keys. The Labconco 
Pressure Controller allows the glove box Main Chamber pressure levels to be set to +/- 5 inches 
of water column [+/- 12.6 mBar]. Once the pressure limits and number of cycles have been 
selected, press the key adjacent to “Start,” and the automated purge and fill process will begin 
and continue for the number of cycles selected.  Each cycle number will be displayed during the 
process. To cancel the automated purge and fill process, press the key adjacent to “Stop” and the 
process will be stopped. 
Note that when the “Glove Box Purge/Fill” mode is in operation, there are no keys available to 
allow for system setup as the “Cycles” key replaced the “SETUP” key. Use the “Mode” key in 
the upper left hand corner adjacent to the icon to return to the “Glove Box Pressure Control” 
screen to use the “Select” mode. 
Manual 
Purge 
Manual 
Fill 
Start/Stop 
Cycles 
Increase 
Upper Limit 
Decrease 
Upper Limit 
Mode Key 
Mute/Alarm 
Increase 
Lower Limit 
Decrease 
Lower Limit 