Instruction Manual
INSTALLATION 
11 
Location 
  Glove box should be placed on a stable, level base stand near an appropriate electrical 
receptacle. The base stand must be capable of supporting 500 lbs minimum, and the height should 
provide a comfortable working position through glove ports (normally 35" to 37" height for 
standing position). 
Connections, Exhaust 
  NOTE: It is the user’s responsibility to determine if the exhaust should be vented to 
outside the building. If it is ducted to the outside, duct restrictions should not exceed 0.5 inches 
water gauge static pressure at 50 CFM. 
Glove Attachment 
  With glove thumbs up and right/left orientation, secure the gloves in place on the glove 
ports by stretching the beaded glove cuff into the groove nearest the window. Install the 8" 
diameter O-ring onto the outer groove over the glove surface. 
Figure 4 
Glove Replacement 
  If it becomes necessary to change gloves without disturbing the integrity of the glove box 
environment, follow these steps: 
1.  Remove the O-ring from the old glove. 
2.  Roll the old glove cuff bead from the inner groove nearest the window to the outside 
groove. 
3.  Insert the new glove through the old glove until the fingers of both are inside the glove box. 
4.  Stretch the new glove’s beaded cuff over the old glove cuff and into the groove nearest the 
window. 
5.  To remove the old glove, grasp the new glove surface and manipulate the old glove bead 
free from the outer groove. 
6.  With the other glove hand inside the box, pull the old glove into the glove box. 
7.  Install the O-ring over the new glove onto the outer groove. 