User’s Manual PuriCare™ Bedding Disposal Stations Models 3840000 3840020 To receive important product updates, complete your product registration card online at register.labconco.com Labconco Corporation 8811 Prospect Avenue Kansas City, MO 64132-2696 800-821-5525, 816-333-8811 FAX 816-363-0130 E-MAIL labconco@labconco.com HOME PAGE www.labconco.com Please read the User’s Manual before operating the equipment.
Copyright © 2003, 2007 Labconco Corporation. All rights reserved. The information contained in this manual and the accompanying products are copyrighted and all rights reserved by Labconco Corporation. Labconco Corporation reserves the right to make periodic design changes without obligation to notify any person or entity of such change. Warranty Labconco provides a warranty on all parts and factory workmanship.
TABLE OF CONTENTS CHAPTER 1: INTRODUCTION About This Manual Typographical Conventions Your Next Step 1 2 3 4 CHAPTER 2: PREREQUISITES Location Requirements Electrical Requirements Space Requirements Overhead Clearance Your Next Step 5 6 6 7 7 7 CHAPTER 3: GETTING STARTED Unpacking Your Bedding Disposal Station PuriCare Components Removing the Shipping Skid Measuring the Inflow Velocity Initial Certification Your Next Step 9 10 11 11 13 13 13 CHAPTER 4: THEORY OF OPERATION AND SAFETY PRECAUTIONS Prefil
CHAPTER 5: USING YOUR PURICARE Starting the Station Reading the Filter System Indicator Working in Your PuriCare Your Next Step 23 23 23 24 25 CHAPTER 6: MAINTAINING YOUR PURICARE Routine Maintenance Schedule Service Operations Speed Control Adjustment Filter Indicator System Adjustment HEPA Filter Replacement HEPA Filter Leak Test Procedure Inflow Velocity Test Procedure Airflow Smoke Patterns Motorized Impeller Replacement Storage 27 28 28 31 32 33 35 37 38 38 39 CHAPTER 7: TROUBLESHOOTING 41 APPEND
CHAPTER 1 INTRODUCTION Congratulations on your purchase of a Labconco PuriCare Bedding Disposal Station. Your PuriCare product is designed to protect you and your laboratory environment from aerosols. It is the result of Labconco’s years of experience manufacturing laboratory equipment, and many of its features were suggested to us by users like you. The PuriCare offers many unique features to enhance safety, performance and ergonomics.
Chapter 1: Introduction About This Manual This manual is designed to help you learn how to install, use, and maintain your bedding disposal station. Chapter 1: Introduction provides a brief overview of the bedding disposal station, explains the organization of the manual, and defines the typographical conventions used in the manual. Chapter 2: Prerequisites explains what you need to do to prepare your site before you install your disposal station. Electrical and service requirements are discussed.
Chapter 1: Introduction Appendix B: Bedding Disposal Station Dimensions contains comprehensive diagrams showing all of the dimensions for the PuriCare model. Appendix C: Bedding Disposal Station Specifications contains the electrical requirements for the Bedding Disposal Station. Wiring diagrams for both the 115V and 230V models are also included. Appendix D: Quick Chart for the PuriCare Bedding Disposal Station provides useful operating specifications.
Chapter 1: Introduction • ) • Critical information is presented in boldface type in paragraphs that are preceded by the wrench icon. These operations should only be performed by a trained certifier or contractor. Failure to comply with the information following a wrench icon may result in injury to the user or permanent damage to your bedding disposal station. Important information is presented in capitalized type in paragraphs that are preceded by the pointer icon.
CHAPTER 2 PREREQUISITES Before you install your Bedding Disposal Station, you need to prepare your site for installation. Carefully examine the location where you intend to install your station. You must be certain that the area is level and of solid construction. In addition, a dedicated source of electrical power must be located near the installation site. Carefully read this chapter to learn: • • • the location requirements for your installation site.
Chapter 2: Prerequisites Location Requirements ! The PuriCare should be located away from traffic patterns, doors, fans, ventilation registers, fume hoods and any other air-handling device that could disrupt its airflow patterns. All windows in the room should be closed. Figure 2-1 shows the optimum location for the PuriCare.
Chapter 2: Prerequisites All PuriCare Stations with model numbers ending in – 00 to –09 are designed for operation at 115 volts, 60Hz, alternating current. PuriCare Stations with model numbers ending in –20 to –29 are designed for operation at 230 +/- 20 volts, 50 Hz alternating current. A dedicated outlet with a circuit breaker rated at 10 amps should be located as close as possible to the right rear side of the station.
Chapter 2: Prerequisites 8 Product Service 1-800-522-7658
CHAPTER 3 GETTING STARTED Now that the site for your bedding disposal station is properly prepared, you are ready to unpack, inspect, install, and certify it. Read this chapter to learn how to: • • • • unpack and move your PuriCare. set up the station. connect the electrical supply source. arrange certification. Depending upon which model you are installing, you may need common plumbing and electrical installation tools in addition to a 3/4" wrench, a flat-blade screwdriver, and a Phillips screwdriver.
Chapter 3: Getting Started Unpacking Your Bedding Disposal Station The United States Interstate Commerce Commission rules require that claims be filed with the delivery carrier within fifteen (15) days of delivery. Carefully unpack your PuriCare and inspect it for damage that may have occurred in transit. If your unit is damaged, notify the delivery carrier immediately and retain the entire shipment intact for inspection by the carrier. ) DO NOT RETURN GOODS WITHOUT THE PRIOR AUTHORIZATION OF LABCONCO.
Chapter 3: Getting Started PuriCare Components Labconco manufactures bedding stations for operation on 115V or 230V. Locate the station model you received in the following group of tables. Verify that the components listed are present and undamaged.
Chapter 3: Getting Started Before Turning on the Unit: 1. Remove the postfilter from its holder on top of the station by pulling it forward, as shown in Figure 3-3. 2. Remove the clear plastic wrapping off of the filter. 3. Reinstall the postfilter, orienting it as shown by the flow arrows on the filter frame. Push the filter back into position until it stops.
Chapter 3: Getting Started Measuring the Inflow Velocity Inflow values and measurement procedures for the PuriCare are described fully in Chapter 6: Maintaining Your PuriCare. Initial Certification Prior to use, all PuriCare stations should be certified by a qualified certifier. Under normal operating conditions, the PuriCare stations should be recertified at least annually and when serviced.
Chapter 3: Getting Started 14 Product Service 1-800-522-7658
CHAPTER 4 THEORY OF OPERATION AND SAFETY PRECAUTIONS The Bedding Disposal Station operates using the following principles: • • Filtration and retention of particulates by High Efficiency Particulate Air (HEPA) filter Directional airflow The major components in the Bedding Disposal Station are: • • • • • • The prefilter The HEPA filter The activated carbon/alumina postfilter The impellers to force air through the unit A speed control for the motor Station air intakes (baffles) and ductwork Product Servic
Chapter 4: Theory Of Operation And Safety Precautions Prefilter The prefilter is located just below the HEPA filter. Its purpose is to lengthen the operating life of the HEPA filter by removing large particles before they can block it. ) REGULARLY SCHEDULED INSPECTION AND REPLACEMENT OF THE PREFILTER IS KEY TO MAINTAINING PROPER OPERATION OF THE ENCLOSURE. HEPA Filter HEPA filters are disposable, dry-type particulate filters.
Chapter 4: Theory Of Operation And Safety Precautions Figure 4-1 The Activated Carbon/Alumina Postfilter The postfilter in your Bedding Disposal Station is unique. Inside of its polyester media are granular activated carbon for the adsorption of odors, plus activated alumina impregnated with potassium permanganate, for additional removal of odors and light gases.
Chapter 4: Theory Of Operation And Safety Precautions Directional Airflow Directional airflow also plays a key role in the Bedding Disposal Station’s performance. Air is drawn into the front of the station and odors. This “curtain” of air makes it difficult for aerosols and odors to escape out of the work area and into the outside environment. This airflow is often calculated and referred to as the Average Inflow Velocity. This is shown graphically in Figure 4-2.
Chapter 4: Theory Of Operation And Safety Precautions Impellers The impellers pull air into the front of the station, through the pre- and HEPA filters, and then discharge it through the postfilter. Speed Control ! The speed control should only be adjusted by a qualified certifier. The speed control is an electronic circuit that allows the certifier to set the impellers’ speed by adjusting its voltage.
Chapter 4: Theory Of Operation And Safety Precautions Safety Precautions The PuriCare Bedding Disposal Station is NOT a biosafety cabinet. Do not use this device for the containment of biohazardous, toxic, flammable, or explosive materials. If you have questions regarding the operation of this device, contact Labconco at 800-821-5525 or www.labconco.com. The PuriCare Bedding Disposal Station should be certified by a qualified certification technician before its initial use.
Chapter 4: Theory Of Operation And Safety Precautions without first disconnecting the PuriCare from electrical service. Avoid the use of flammable gases or solvents in the PuriCare. Care must be taken to ensure against the concentration of flammable or explosive gases or vapors. An open flame should NOT be used in the PuriCare. Open flames may disrupt the airflow patterns in the station or damage the filters’ adhesive.
Chapter 4: Theory Of Operation And Safety Precautions ! ) Proper operation of the station depends largely upon its location and the operator’s work habits. Consult the Installation and Normal Operation sections of this manual for further details. WHEN SURFACE DISINFECTING THE PURICARE STATION: • • • AVOID SPLASHING THE DISINFECTING SOLUTION ON SKIN OR CLOTHING. ENSURE ADEQUATE VENTILATION.
CHAPTER 5 USING YOUR PURICARE Starting the Station To start the PuriCare, turn the blower switch to the “ON” position, as shown in Figure 5-1. Reading the Filter System Indicator The two light system located on the control panel as shown in Figure 5-1 indicates the total system pressure. During operation, a green “NORMAL” indicator light stays on. When the pre- or HEPA filters become excessively loaded, the “NORMAL” light turns off and the “SERVICE” light turns on, indicating the unit should be serviced.
Chapter 5: Using Your PuriCare Figure 5-1 Working In Your PuriCare Planning • Thoroughly understand procedures and equipment required before beginning work. • Arrange for minimal disruptions, such as room traffic or entry into the room while the station is in use. Start-up • Turn on fluorescent light and station blower. • Check the airfoil and baffle for obstructions, and note the indicator lights. • Allow the station to operate unobstructed for 5 minutes.
Chapter 5: Using Your PuriCare Bedding Disposal Operations • Do not obstruct the front airfoils or rear baffle. • Large objects should not be placed close together. • Keep all materials at least 4 inches inside of the airfoil, and perform all operations as far to the rear of the work area as possible. • Avoid the use of an open flame. • Avoid using techniques or procedures that disrupt the airflow patterns of the station.
Chapter 5: Using Your PuriCare 26 Product Service 1-800-522-7658
CHAPTER 6 MAINTAINING YOUR PURICARE Now that you have an understanding of how to work in the Bedding Disposal Station, we will review the suggested maintenance schedule and the common service operations necessary to maintain your PuriCare for peak performance. ! Many of the service operations should be performed only by trained and experienced certification technicians after the station has been properly decontaminated. DO NOT attempt to perform these operations if you are not properly trained.
Chapter 6: Maintaining Your PuriCare Routine Maintenance Schedule After Each Work Session • Inspect the prefilter and replace if needed. Weekly • Using 70% ethanol, or a suitable disinfectant, surface disinfect the inside of the station, and the work surface. • Operate the station blower, noting the indicator lights. Monthly (or more often as required) • Using a damp cloth, clean the exterior surfaces of the station, particularly the front and top of the unit, to remove any accumulated dust.
Chapter 6: Maintaining Your PuriCare Prefilter replacement: 1. Turn the station off. 2. Loosen the four thumbscrews that secure the prefilter access panel to the top of the work area, as shown in Figure 6-1. Let the access panel swing down. 3. Pull the prefilter straight out and down. 4. Install a new prefilter, orienting the prefilter airflow directional arrows up towards the HEPA filter. Figure 6-1 Postfilter replacement: 1.
Chapter 6: Maintaining Your PuriCare 4. Install the postfilter, orienting it as shown by the flow arrows on the filter frame. Push the filter back into position until it stops. Figure 6-2 Dress Panel Removal: 1. Locate the front dress panel, just above the control panel. Remove the two screws located on the top of the panel. Remove the panel by pulling it slightly upward and out, as shown in Figure 6-3.
Chapter 6: Maintaining Your PuriCare Figure 6-3 Fluorescent Lamp Removal: 1. Remove the front dress panel as described above. 2. Locate the control panel on the front of the cabinet, which has the filter indicator lights and switches. Remove the screws located on each lower end of the panel. Swing the panel upward for access to the fluorescent lamp, as shown in Figure 6-3. 3. Remove the fluorescent lamp by rotating the lamp and pulling it straight out of its sockets. 4.
Chapter 6: Maintaining Your PuriCare 3. Replace the front dress panel and reestablish the inflow air velocity. Figure 6-4 Filter Indicator System Adjustment ! The Filter Indicator System was adjusted at the factory and should not require recalibration until the HEPA filter is replaced. To adjust system, do the following: 1. Ensure that the prefilter is new and the station is operating within its airflow tolerance. 2.
Chapter 6: Maintaining Your PuriCare 5. If the ‘NORMAL’ light is on, remove the dress panel, and using the appropriate hex wrench, turn the adjustment switch slowly clockwise as shown in Figure 6-5. Reinstall the dress panel. Repeat this procedure until the ‘SERVICE’ light turns on when the dress panel is replaced. 6. If the ‘SERVICE’ light is on, remove the dress panel, and using the appropriate hex wrench, turn the adjustment switch slowly counterclockwise as shown in Figure 6-5.
Chapter 6: Maintaining Your PuriCare 3. Remove the two access hole plugs on the top of the unit. 4. Using the 9/16” deep socket, loosen the filter clamp nuts located in the access holes. Refer to Figure 6-6 for further details. 5. With the clamp bolts loosened, the supply filterclamping frame should be clear of filter. Carefully pull the filter straight out of the unit and discard properly. 6. With the filter removed, inspect the clamping frame and the frame of the cabinet for damage. 7.
Chapter 6: Maintaining Your PuriCare Figure 6-6 HEPA Filter Leak Test Procedure To ensure dirty room air does not affect the HEPA Filter Leak Test and that the control panel and sash gaskets are sealing, take the following steps: 1. Turn on the enclosure and set the average inflow velocity as indicated in inflow velocity test procedure. Remove the postfilter 2. Tape closed all openings full length across top, front and sides of the front dress panel as shown in Figure 6-7. 3.
Chapter 6: Maintaining Your PuriCare 4. Place photometer pickup over the exhaust discharge and sample the concentration. The concentration must not exceed 0.005%. 5. If the unit fails, check the gasket between the control panel and the upper cabinet flange. Also, check gasket attached to underside of the upper cabinet for contact with the upper edge of the sash when it is in the closed position. 6. Use a Laskin nozzle-type generator. DO NOT use more than one nozzle. Operate the generator at 10 PSIG.
Chapter 6: Maintaining Your PuriCare Inflow Velocity Test Procedure Inflow velocity can be determined by use of a thermal anemometer at the face of the work access opening by taking the following steps: 1. If the meter has an adjustable time constant, set it at 10 to 15 seconds. 2. Take three rows of readings at 8, 12 and 16 inches from the top of the opening at positions that are 6 and then 12 inches in from both sidewalls. This will yield a grid of 12 sampling points. See Figure 6-8 for probe placement.
Chapter 6: Maintaining Your PuriCare Airflow Smoke Patterns Non-thermal smoke should be released from one side of the enclosure to the other, along a line in the middle of the work area, two inches above the work surface. The smoke should flow directly into the rear baffle. Smoke released near the front end of the enclosure, or at the top of the enclosure may exhibit vortexing and a reflux along the sidewalls or top of the enclosure.
Chapter 6: Maintaining Your PuriCare 4. Replace the capacitor with a new one of equal voltage and capacity. 5. Reassemble the unit by reversing the disassembly steps. Recertify the enclosure. Storage If the Bedding Disposal Station is to be left unused for more than one month, the unit should be prepared for storage. 1. Surface disinfect or contaminate the cabinet as required. 2. Unplug the unit. 3. Cover and seal the work area access opening and the exhaust outlet with plastic sheeting. 4.
Chapter 6: Maintaining Your PuriCare 40 Product Service 1-800-522-7658
CHAPTER 7 TROUBLESHOOTING Refer to the following table if your Bedding Disposal Station fails to operate properly. If the suggested corrective actions do not solve your problem, contact Labconco for additional assistance. PROBLEM CAUSE CORRECTIVE ACTION Station impellers and lights won’t turn on Unit not plugged into outlet Plug the PuriCare into appropriate electrical service. Circuit breakers tripped Reset circuit breakers.
Chapter 7: Troubleshooting PROBLEM CAUSE CORRECTIVE ACTION Cabinet impeller turns on but lights don’t work (cont.) Lamp is defective Inspect lamp wiring.
APPENDIX A BEDDING DISPOSAL STATION COMPONENTS Illustration A-1 indicates the location of the following service parts: Bedding Disposal Station Replacement Parts Item Quantity Part No.
Appendix A: PuriCare Components A-1 44 Product Service 1-800-522-7658
APPENDIX B BEDDING DISPOSAL STATION DIMENSIONS B-1 Product Service 1-800-522-7658 45
Appendix B: Bedding Disposal Station Dimensions 46 Product Service 1-800-522-7658
APPENDIX C BEDDING DISPOSAL STATION SPECIFICATIONS Electrical Data Station Model 3840000 to –09 Electrical Requirements 115 VAC – 60 Hz, 1 Phase – 5 Amps Station Model 3840020 to –29 Electrical Requirements 230 VAC – 50/60 Hz, 1 Phase – 3 Amps Impeller Specifications (2 each) Station Model 3840000 to –09 Electrical Requirements 115 VAC – 60 Hz, 100 Watts, 2550 RPM, Automatic Thermal Protection Station Model 3840020 to –29 Electrical Requirements 230 VAC – 50/60 Hz, 100 Watts, 2550 RPM, Automatic Ther
Appendix C: Bedding Disposal Station Specifications Environmental Conditions • • • • • • • 48 Indoor use only. Maximum altitude: 6562 feet (2000 meters). Ambient temperature range: 41° to 104°F (5° to 40°C). Maximum relative humidity: 80% for temperatures up to 88°F (31°C), decreasing linearly to 50% relative humidity at 104°F (40°C). Main supply voltage fluctuations not to exceed ±10% of the nominal voltage.
Appendix C: Bedding Disposal Station Specifications C-1 Product Service 1-800-522-7658 49
Appendix C: Bedding Disposal Station Specifications 50 Product Service 1-800-522-7658
APPENDIX D QUICK CHART FOR THE PURICARE BEDDING DISPOSAL STATONS Model Opening Height (inches) Inflow (FPM) Work Area (ft2) Total Air Volume Displacement (CFM) Number of Laskin Nozzles Needed HEPA Filter Dims. (in.) Fluorescent Light(s) Product Service 1-800-522-7658 38400 23 65 or greater 5.
Appendix D: Quick Chart for the PuriCare Bedding Disposal Station 52 Product Service 1-800-522-7658
APPENDIX E REFERENCES Many excellent reference texts and booklets are currently available. The following is a brief listing: Primary Containment for Biohazards: Selection, Installation and Use of Biological Safety Stations. September 1995. U.S. Department of Health and Human Services, Public Health Service, Centers for Disease Control and Prevention and National Institutes of Health. Biosafety in Microbiological and Biomedical Laboratories. 4th Edition, May 1999. U.S.
Appendix E: References 54 Product Service 1-800-522-7658
DECLARATION OF CONFORMITY Application Council Directive(s): 73/23/EEC, 89/336/EEC Standard(s) to which conformity is declared: EN61010, EN55022, EN50082-1 Manufacturer’s Name: Labconco Corporation Manufacturer’s Address: 8811 Prospect Avenue Kansas City, MO 64132 USA Importer’s Name: See Shipping/Customs Documents* Importer’s Address: See Shipping/Customs Documents for your equipment Type of Equipment: Laboratory Equipment Model No: Serial No.