Chapter 3: Getting Started
Measuring the Inflow
Velocity
Inflow values and measurement
procedures for the PuriCare are
described fully in Chapter 6:
Maintaining Your PuriCare.
Initial Certification
Prior to use, all PuriCare stations should be certified by
a qualified certifier. Under normal operating
conditions, the PuriCare stations should be recertified at
least annually and when serviced. The certifier should
perform the following tests:
HEPA Filter Leak Test
Inflow Velocity Test
Airflow Smoke Patterns
If you have any questions regarding certification
agencies or need assistance in locating one, contact
Labconco’s Product Service Department at 1-800-522-
7658 or 816-333-8811.
Your Next Step
After your PuriCare has been installed and certified,
you are ready to proceed to Chapter 4: Theory Of