Owner's manual
Page 132 of 197 
Figure 8-16 
Blower Mounting Bolts 
The motor/blower assembly is heavy. Handle with care. 
7.  While supporting the blower assembly, pull the assembly straight out the plenum, as 
shown in Figure 8-17. 
Figure 8-17 
8.  To replace the motor/blower assembly, reverse the above procedure. 
Note: During reinstallation of the blower assembly, angle the rear blower 
bracket into the plenum blower opening. Make sure to engage the rear 
blower bracket inside the plenum blower hole. After reinstalling and 
tightening the three blower bolts, ensure that the rear blower bracket is 
correctly engaged by trying to move the blower assembly. If it is rigidly 
captured on the plenum, then the rear bracket is properly engaged. If the 
rear of the blower assembly moves freely, then the blower must be 
reinstalled. 
9.  Continue the reassembly. Plug in the cabinet and recertify it before use. 