5.  Remove the two Phillips screws on the front edge of the right side corner post, 
and allow the front of the side panel to swing out and lift off.     
6.  Continue routing the sensor cable 
along the main harness on the right 
side of the cabinet. Secure the 
sensor harness with the remaining 
ties as shown.     
7.  Route the cable to avoid pinch hazards, and to bring the two airflow sensor 
connectors to the control board. Insert the two connectors into the control 
board. One harness connector has 5 pins, and the other has 6. Connect each to 
the corresponding pin connectors on the board as shown in the following photo. 
Airflow Sensor connectors 