Owner's manual
Page 165 of 197 
Canopy Air Sensor Kit No. 3896600 
Note: This kit is designed to install a digital (Inflow and Downflow) air sensor 
in a canopy-connected Type A2 Logic only. If you are want to install a digital 
airflow sensor in a non-canopy connected Logic, then you must use kit # 
3859400.   
Note: A qualified certifier must install and calibrate this sensor to obtain 
accurate operation. 
Installation Procedure 
1.  Disconnect power to the Biosafety Cabinet. 
2.  Open the right side panel of the Logic by removing the two Phillips screws on the 
front edge of the panel. 
3.  Let the panel swing out, and examine the control board on the side of the 
cabinet as shown. 
4.  Note the orientation of the ribbon and 
touchpad connectors before 
disconnecting them from the board. 
5.  Loosen the bottom nut, and remove 
the top nut to remove the control 
circuit board assembly from the 
chassis of the Logic 
6.  Note the orientation of the control 
circuit board in the metal frame. 
Disassemble the four screws and nuts 
securing the control circuit board to 
the metal frame as shown.   