Owner's manual
Page 93 of 197 
Inflow Velocity Calculation - Secondary Method 
If the primary method cannot be performed, the secondary inflow calculation method 
should be used. The secondary method uses a hot wire thermal anemometer. In this 
method, the sash is lowered, and a series of velocity readings are taken at the front of 
the unit. 
NOTE: You must use Labconco holder P/N 3836405 to perform this test 
properly. Qualified certifiers can obtain this holder as part of the Logic 
Certifier Kit. 
For each model Purifier, the sash should be closed until the distance from the bottom 
edge of the sash to the mating surface of the grey corner post end cap, equals 
“Restricted Sash Opening” as shown Table 6-5 below. The measurement should be 
taken using the appropriate marked template as shown below in Figure 6-9 below. 
The thermal anemometer probe should be positioned such that the distance from the 
bottom of the probe holder to the center of the sensor element equals “Sensor Distance” 
as shown in Table 6-5 below. Use the marks on the 6-foot templates, or a scale as 
shown in Figure 6-10a & b below. 
Figure 6-9 
Measuring the restricted access 
opening height, using the 
appropriate template. Ensure that 
the bottom of the template is 
seated in the sash stop. 










