Drain Valve Installation 
In order to prevent damage during shipping, the drain valve assembly has not 
been installed. If desired, the valve should be installed after the cabinet is in 
its final location. 
To install the valve assembly, follow these steps: 
Note: The work surface is heavy. Use caution when handling it. 
4.  Lift the work surface out of the Procedure Station by lifting on the knobs 
at the front of the work surface. Steady the work surface while pulling it 
straight out the front of the cabinet. 
5.  Using a putty knife, remove and discard the stainless steel cover that is 
sealed over the drain mounting holes. Scrape out remaining sealant that is 
around the holes. 
6.  Apply a light coating of silicone sealant (user supplied) to the mounting 
surface of the drain assembly. Attach the drain assembly under the 
bottom of the cabinet as shown in Figure 3-3. Wipe off any excess sealant 
from the cabinet bottom. Ensure that the center drain hole is unobstructed. 
7.  Make sure the drain valve is in the closed position. 
8.  Reinstall the work surface. 
9.  Allow the silicone sealant to cure for at least eight hours before exposing 
it to liquid. 
NOTE: The drain valve assembly 
attaches to the underside of the cabinet 
bottom. 
Apply a light coat of silicone sealant to 
this surface of the connector, aligning the 
three holes in the connector with the three 
holes in the procedure station liner. 
Figure 3-3 