Motor/Blower 
The motor/blower assembly pulls room air into the front of the cabinet, and 
re-circulates it internally. During its recirculation, the air is split into two 
separate streams. One path leads through the exhaust HEPA filter and out of 
the unit. The second path flows through the supply HEPA filter, which then 
flows down through the work area, as shown in Figure 4-4. The motor in the 
PuriCare station is an electronically commutated motor (ECM). The ECM is 
a brushless DC motor that includes its own power supply to convert the 
incoming alternating current to direct current, as well as its own 
microprocessor to control and measure the motor’s operation. The motor 
utilizes Labconco’s exclusive Constant Airflow Profile
TM
 (CAP) program to 
deliver a consistent volume of air, throughout the life of the cabinet. 